MSM07 Series
7W Medical Open Frame Power Supply
- Wide Range AC Input
- Fully Regulated DC Output
- Lifetime Expectation >5 years
- Hold-up Time >22ms at full load
- Output Floating
- Safety Approval - EN60601-1 Class II


Print DatasheetModel Overview Input/Output Configuration Mechanical Drawings Supporting Documentation Ordering Options Technical Support AvailabilityElpac Part Number Output Voltage Output Current Peak Current1 Total Regulation2 Typical Efficiency3 MSM0705 5.0V 1.4A 2.0A ±5% 74% MSM0712 12.0V 0.6A 0.83A ±5% 74% MSM0715 15.0V 0.5A 0.66A ±5% 76% MSM0724 24.0V 0.3A 0.42A ±5% 75% MSM0728 28.0V 0.3A 0.36A ±5% 75% Notes
1 Maximum peak load (10W) lasting 500ms with a maximum 10% duty cycle.
2 Includes initial setting, line regulation, load regulation, and thermal drift.
3 Typical at 115VAC and full load.
Input Input Voltage 85 - 264VAC 100 - 240VAC Nominal Input Frequency 50 - 60Hz Input Current <0.3A rms Inrush Current <37A at 230VAC cold start Zero Load Power Consumption <0.7W Earth Leakage Current (Typical) <32µA @ 132VAC @ 60Hz <55µA @ 264VAC @ 60Hz Output Output Voltage See Table Total Regulation +/-5% Minimum Load No minimum load required Start-Up Delay <750 ms Hold-Up Time >22ms Ripple & Noise <1% pk-pk ** * Over Voltage Protection 110-135% Over Current Protetion 200 - 350% Short Circuit Protection shutdown, auto-restart (hiccup mode) Notes
* Ripple and noise measured with 20MHz bandwidth; 10μF tantalum capacitor in parallel with a 0.1μF ceramic capacitor.
General Efficiency Avg Efficiency 74.8% @ 115VAC; 75.9% @ 230VAC MTBF min. 200,000 hours demonstrated Size 2.00" x 2.28" x 0.71" | 50.80mm x 57.91mm x (18.11mm Weight 0.10 lbs (0.05 kg) Power Density 2.2W/in3 Environmental Operating Temperature 0 – 50°C (Full load to 50°C, derate linearly to 50% load at 70°C) Storage Temperature -40°C to +85°C Relative Humidity 5-95%, non-condensing Cooling Natural Convection Vibration All units production tested to 19.6m/s2 EMC & Safety Emissions FCC Part 15, Class B, CISPR11, EN61000-3-2, -3 Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11 Certified by UL to the following: 
cЯUus UL 60601-1 CAN/CSA-22.2 No.601.1-M90 CB per IEC60601-1 CE marked to LVD Input Configuration Connector Molex p/n 22-43-8030 Mating Connector Molex p/n 50-37-5033 or Equivalent Mating Connector Contacts Molex p/n 08-70-1040 or Equivalent Output Configuration Connector Molex p/n 22-43-8020 Mating Connector Molex p/n 50-37-5023 or Equivalent Mating Connector Contacts Molex p/n 08-70-1040 or Equivalent Input Pin Assignments (P1) Pin 1 AC Line Pin 2 <not assembled> Pin 3 AC Neutral Output Pin Assignments (P2) Pin 1 +V1 Pin 2 Com

ICCNexergy has the capability to modify our standard power supplies to meet custom specifications. With a broad standard power platform, we can quickly deliver a modified power solution in as little as three weeks. Modified options include:- Grounded Output
- Cooling
- Chassis Mount
- Cable Harness
- Load Share
Inventory Results For MSM07
Part Number Distributor Quantity Inv. Date MSM0705 NEWARK/ELEMENT14 21 6/18/2013 MSM0705 PEI-GENESIS 4 6/18/2013 MSM0705 DIGI-KEY CORPORATION 163 6/15/2013 MSM0712 NEWARK/ELEMENT14 23 6/18/2013 MSM0712 PEI-GENESIS 1 6/18/2013 MSM0712 ALLIED ELECTRONICS, INC. 7 6/17/2013 MSM0712 DIGI-KEY CORPORATION 249 6/15/2013 MSM0712 MOUSER ELECTRONICS 47 6/15/2013 MSM0715 NEWARK/ELEMENT14 3 6/18/2013 MSM0715 PEI-GENESIS 3 6/18/2013 MSM0728 NEWARK/ELEMENT14 3 6/18/2013 MSM0728 PEI-GENESIS 3 6/18/2013 Search Part Database